WARNING LETTER
•ACell, Inc•December 20, 2012

ACell, Inc – FDA Warning Letter, Dec 20, 2012

Record Details

On April 26, 2013, the FDA issued a Warning Letter to ACell, Inc. following an inspection from November 28 to December 20, 2012, at their Columbia, Maryland facility. The inspection revealed that ACell's MatriStem® Surgical Matrix Thick (PSMT) devices are adulterated under 21 U.S.C. § 351(h) because their manufacturing, packing, storage, or installation methods do not conform to Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation (21 C.F.R. Part 820).

Key violations include: 1. **Failure to establish and maintain adequate design control procedures (21 CFR § 820.30(a))**: * Lack of rationale for increased design input values (e.g., tensile burst strength). * Incomplete Design History File (DHF) deliverables. * Design validation was not conducted on production equivalents of the device. * The two-year expiration date for Multi-Layer Vacuum Pressed product is not supported by real-time stability data. The firm's use of accelerated aging is deemed inadequate due to the complex nature of the product, requiring real-time data or validated accelerated aging. Stability assessment parameters were also found insufficient, lacking specifications for onset temperature and hydration, and

Company
ACell, Inc
Inspection date
December 20, 2012
Product type
Devices
Office
Baltimore District Office
People
  • Evelyn Bonnin (District Director)
  • Rodney W. Bosley
View Company Profile

ID · 0ab58bcc-9829-4937-b5a0-cc96b3f59a2f

Violation Codes8
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(a)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 80321 CFR 803.50(a)(1)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore ACell, Inc
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.