WARNING LETTER
•Cardiac Assist, Inc•February 7, 2014

Cardiac Assist, Inc – FDA Warning Letter, Feb 7, 2014

Record Details

On May 13, 2014, the FDA issued a Warning Letter to Cardiac Assist, Inc. following an inspection from January 24, 2014, to February 7, 2014. The inspection revealed that the firm's TandemHeart PTVA Blood Pump and TandemHeart Escort Controller are misbranded under Section 502(t)(2) of the Act, due to a failure to furnish required material or information under Section 519 and 21 CFR Part 803 - Medical Device Reporting.

A significant violation identified was the failure to submit a Medical Device Report (MDR) to the FDA within 30 calendar days of becoming aware of information suggesting a device may have caused or contributed to a death, as required by 21 CFR 803.50(a)(1). Specifically, the firm was aware of a complaint where a TandemHeart pump stopped turning, and the patient subsequently died, but no MDR was filed.

The FDA acknowledged the firm's February 26, 2014, response to the FDA-483 appeared sufficient for Quality System deficiencies, but emphasized adherence to procedures and regulations. The letter mandated prompt corrective action to prevent regulatory actions like seizure, injunction, or civil money penalties. Cardiac Assist, Inc. was required to notify the FDA in writing within fifteen business days of the specific steps taken to correct

Company
Cardiac Assist, Inc
Inspection date
February 7, 2014
Product type
Devices
Office
Philadelphia District Office
Person
  • Anne E. Johnson (Senior Science Advisor)
View Company Profile

ID · 887fb125-9267-48db-af8c-849b5d1a988e

Violation Codes5
21 U.S.C. 321(h)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 80321 CFR 803.50(a)(1)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Cardiac Assist, Inc
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.