Attention www.globalkeysolutions.net is now www.keypedia.com, read the full press release here.
KeyPedia LogoKeyPedia Icon
  1. Home
  2. /
  3. Records
  4. /
  5. Warning Letter
  6. /
  7. CareFusion 303, Inc.
WARNING LETTER
•CareFusion 303, Inc.•May 22, 2024

CareFusion 303, Inc. – FDA Warning Letter, May 22, 2024

Record Details

The FDA issued a Warning Letter to the firm following a May 6-22, 2024, inspection in San Diego, California, concerning its Pyxis Medication Management System medical devices. The devices were found adulterated under section 501(h) due to non-conformity with Quality System regulation (21 CFR Part 820), and the firm's responses were deemed inadequate. Key violations include the failure to establish and implement corrective and preventive action (CAPA) procedures (21 CFR 820.100(a)), evidenced by numerous open, severe software defect tickets. The firm also failed to establish adequate complaint handling procedures (21 CFR 820.198(a)), with thousands of complaints lacking proper investigation or MDR evaluation, including those related to patient harm and electrical safety. Design validation was deficient, lacking adequate risk analysis (21 CFR 820.30(g)) for high-severity risks like patient death and cybersecurity vulnerabilities. Furthermore, the firm failed to submit Medical Device Reports (MDRs) within 30 days for malfunctions likely to cause death or serious injury (21 CFR 803.50(a)(2)), with multiple instances of delayed reporting. Lastly, the firm failed to submit Reports of Correction or Removal within 10 working days for actions addressing health risks (21 CFR § 806.10), such as product advisories and software updates. The FDA requires a written response within fifteen business days detailing specific corrective actions, a timetable, and evidence of implementation. Failure to address these violations may result in regulatory actions, including seizure, injunction, civil money penalties, and impact on federal contracts, device approvals, and Certificates to Foreign Governments.

Company
CareFusion 303, Inc.
Inspection date
May 22, 2024
Product type
Devices
Office
Center for Devices and Radiological Health
People
  • Sean M. Boyd (Office Director)
  • Raymond W. Brullo
  • Connor L. Bates
  • Vipul Sheth
  • Jeffrey R. Wooley (Compliance Officer)
View Company Profile

ID · ac8b65c0-fcdf-4886-b883-966cf930161a

Violation Codes9
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 820.100(a)21 CFR 820.198(a)21 CFR 820.30(g)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 803.50(a)(2)21 CFR 806.10

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore CareFusion 303, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.