Childrens Medical Ventures – FDA Warning Letter, Mar 28, 2014

Record Details

The FDA issued a Warning Letter to Children’s Medical Ventures/Philips Healthcare following an inspection from March 10-28, 2014, at their former Monroeville, PA, facility, which manufactures the SmartMonitor 2 PS device. The device was deemed adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System regulation (21 CFR Part 820) and misbranded under 21 USC 352 (t)(2) for failing to furnish required information.

Violations include: 1. **Failure to control nonconforming product (21 CFR 820.90(a)):** A wiring issue in 311 SmartMonitor 2 PS battery packs (lots (b)(4) and (b)(4)) caused device failures, constant alarms, and disrupted monitoring of infant respiration, heart rate, and oxygen saturation. No Health Hazard Evaluation or risk reduction activities were performed. As of March 13, 2014, 99 of 143 affected units distributed to the field remained uncontrolled. 2. **Inadequate complaint file maintenance (21 CFR 820.198(a)):** Twenty-seven service work orders related to Field Change Order (FCO) reference (b)(4) were not maintained as complaint files. 3.

Company
Childrens Medical Ventures
Inspection date
March 28, 2014
Product type
Devices
Office
Philadelphia District Office
Person
  • Anne E. Johnson (Senior Science Advisor)
View Company Profile

ID · 831eb08b-8b5e-4c68-af0c-bf91ce694348

Violation Codes8
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.90(a)21 CFR 820.198(a)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 806.20(b)(5)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Childrens Medical Ventures
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.