WARNING LETTER
•Circulatory Technology, Inc.•September 26, 2018

Circulatory Technology, Inc. – FDA Warning Letter, Sep 26, 2018

Record Details

On January 28, 2019, the FDA issued a Warning Letter to Circulatory Technology, Inc. following an inspection from August 22, 2018, to September 26, 2018, at their Oyster Bay, NY facility. The inspection determined that the firm is a specification developer for the Better Bladder and Bigger Better Bladder medical devices.

The devices were found to be adulterated under section 501(h) of the Act, as their manufacturing, packing, storage, or installation methods did not conform to the Quality System regulation (21 CFR Part 820). The firm's October 22, 2018, response to the FDA 483 was deemed inadequate.

Violations included: 1. **Failure to perform and document revalidation activities (21 CFR 820.75(c))**: The firm added an assembly step in February 2018 to the Better Bladder manufacturing process but did not revalidate it. The firm's argument that a leak test and lack of complaints suffice was rejected. 2. **Failure to establish and maintain design control procedures (21 CFR 820.30(a))**: The February 2018 manufacturing change was not reviewed as part of design control. The firm's assertion that it wasn't a design change was deemed inadequate

Company
Circulatory Technology, Inc.
Inspection date
September 26, 2018
Product type
Devices
Office
New England District Office
Person
  • Joseph S. Matrisciano (Program Division Director)
View Company Profile

ID · 7e92b87d-86ac-44e4-9ee9-1209312dbfbb

Violation Codes7
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.75(c)21 CFR 820.30(a)21 CFR 820.198(a)(3)21 CFR 820.100(b)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Circulatory Technology, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.