Coastal Implant Technologies, Inc. – FDA Warning Letter, Oct 6, 2010
On February 3, 2011, the FDA issued a Warning Letter to Coastal Implant Technologies, Inc. following an inspection from August 31, 2010, to October 6, 2010. The inspection revealed that the firm, identified as a manufacturer (specification developer) of Coastal Implant Technologies brand, Newport (Straight-wall) and Sunset (Cylindrical-wall) endosseous dental implants, was in violation of Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation (21 CFR Part 820).
Key violations included: 1. **Failure to establish and maintain procedures for purchased products and services (21 CFR 820.50):** The firm lacked procedures and records to ensure conformity of contract-manufactured dental implants. Their claim of being only a distributor was deemed inadequate due to the absence of a written agreement with their manufacturer and lack of visual inspection records. 2. **Failure to establish and maintain design control procedures (21 CFR 820.30(a)):** No procedures, records, or files existed for the design and development of their dental implants. The firm's assertion that the manufacturer handled design controls was insufficient without a written purchasing control agreement. 3. **Failure to establish and maintain procedures for nonconforming product control (21 CFR 820.90(a)):**
ID · 9f1081d4-3de8-4a39-aac5-4577739d0d76
Full citation text and observation details available on the Dashboard.
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.