WARNING LETTER
•Cortechs Labs Inc•August 5, 2010

Cortechs Labs Inc – FDA Warning Letter, Aug 5, 2010

Record Details

An FDA inspection of Cortechs Labs Inc. from July 30 to August 05, 2010, revealed that the firm manufactures and distributes picture archiving communications systems (PACS), which are medical devices. The inspection determined that these devices are adulterated under Section 501(h) of the Federal Food, Drug, and Cosmetic Act because the manufacturing methods, facilities, or controls do not conform to the Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation (21 CFR Part 820). Significant violations identified include the lack of adequately established procedures for corrective and preventive actions (21 CFR 820.100(a)), complaint handling (21 CFR 820.198(a)), and document control (21 CFR 820.40). Additionally, the firm failed to perform quality audits (21 CFR 820.22), document management review results (21 CFR 820.20(c)), complete risk analysis (21 CFR 820.30(g)), and establish procedures for purchased products and services (21 CFR 820.50). Cortechs Labs Inc. must promptly correct these violations and respond in writing within fifteen working days, outlining specific corrective steps and providing supporting documentation. Failure to comply may result in enforcement actions such as seizure, injunction, civil money penalties, and non-approval of pre-market applications for Class III devices.

Company
Cortechs Labs Inc
Inspection date
August 5, 2010
Product type
Devices
Office
Department of Health and Human Services
Person
  • Alonza E. Cruse (Director)
View Company Profile

ID · a6895672-4e80-486d-a23e-81cdf45dd473

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.100(a)21 CFR 820.198(a)21 CFR 820.4021 CFR 820.2221 CFR 820.20(c)21 CFR 820.30(g)21 CFR 820.50

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Cortechs Labs Inc
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.