Excel Medical Products, Inc. – FDA Warning Letter, Dec 21, 2012
On April 9, 2013, the FDA issued a Warning Letter to Excel Medical Products, Inc. following an inspection from November 19 to December 21, 2012. The inspection revealed that the firm's Class II Hemostatic Y-connectors are adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations include: 1. **Process Validation (21 CFR 820.75(a)):** Failure to validate injection molding processes for various Y-connector components (e.g., 6 Fr. cap, rotator, Y-body components across different sizes and colors). The firm's response, indicating planned updated validations, was deemed inadequate as it involved future actions. 2. **Design History File (DHF) (21 CFR 820.30(j)):** Absence of DHFs for 4.5, 7, and 10 Fr. Hemostatic Y-connector devices. The firm's plan to compile design files after validation was not yet evaluable. 3. **Nonconforming Product Control (21 CFR 820.90(a)):** Lack of procedures for documenting and evaluating nonconforming product from production set-up. Planned new setup procedures and
ID · a7057dc9-ebf0-4b10-b4c1-e8eb3162f2ba
Full citation text and observation details available on the Dashboard.
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.