Attention www.globalkeysolutions.net is now www.keypedia.com, read the full press release here.
KeyPedia LogoKeyPedia Icon
  1. Home
  2. /
  3. Records
  4. /
  5. Warning Letter
  6. /
  7. Fisioline s.r.l.
WARNING LETTER
•Fisioline s.r.l.•November 1, 2012

Fisioline s.r.l. – FDA Warning Letter, Nov 1, 2012

Record Details

An FDA inspection of Fisioline, s.r.l in Verduno, Italy, from October 29 to November 1, 2012, revealed that the firm's Lumix 2 and Lumix 3 laser devices are adulterated and misbranded. The devices are adulterated under section 501(h) of the Act because manufacturing methods and controls do not conform to the Quality System (QS) regulation (21 CFR Part 820). They are misbranded under section 502(t)(2) for failing to provide information required by the Medical Device Reporting (MDR) regulation (21 CFR Part 803). The firm's response to the FDA 483 was not reviewed due to late submission. Significant QS violations include failures in establishing and maintaining procedures for complaint handling (21 CFR 820.198(a), (b), (e)), computer software validation (21 CFR 820.70(i)), device design validation, verification, and review (21 CFR 820.30(g), (f), (e)), corrective and preventive action (21 CFR 820.100(a)), nonconforming product disposition (21 CFR 820.90(b)(1)), equipment calibration and documentation (21 CFR 820.72(a), (b)(2)), acceptance activities documentation (21 CFR 820.80(e)), and quality audits (21 CFR 820.22). Additionally, the firm failed to develop, maintain, and implement written MDR procedures as required by 21 CFR 803.17. Fisioline must notify the FDA within fifteen business days with specific corrective actions, including systemic changes and a timetable for implementation. Failure to correct these violations may impact federal contracts, premarket approvals for Class III devices, and the granting of Certificates to Foreign Governments.

Company
Fisioline s.r.l.
Inspection date
November 1, 2012
Product type
Devices
Office
Center for Devices and Radiological Health
Person
  • Steven D. Silverman (President)
View Company Profile

ID · a5a328a8-f207-4975-89a8-15b39e7edbca

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.198(a)21 CFR 820.198(b)21 CFR 820.198(e)21 CFR 820.70(i)21 CFR 820.30(g)21 CFR 820.30(f)21 CFR 820.30(e)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Fisioline s.r.l.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.