WARNING LETTER
•Fresenius Medical Care, North America•December 7, 2012

Fresenius Medical Care, North America – FDA Warning Letter, Dec 7, 2012

Record Details

An FDA inspection of Fresenius Medical Care, North America in Ogden, UT, from November 26 to December 7, 2012, revealed that their Optiflux Polysulfone Dialyzers, classified as medical devices, were adulterated. This adulteration stems from non-conformity with the Quality System regulation (21 CFR Part 820), specifically regarding manufacturing controls. The firm's responses dated December 28, 2012, January 2, 2013, and February 7, 2013, were deemed inadequate by the FDA. Key violations include the failure to establish and maintain adequate procedures for verifying device design, as required by 21 CFR 820.30(f). This was evidenced by incomplete biocompatibility testing, lack of documentation for required additional testing, and insufficient design verification studies. Additionally, the firm failed to establish and maintain adequate procedures for validating device design, as mandated by 21 CFR 820.30(g). Deficiencies included incomplete design validation for Optiflux PS dialyzers, a clinical study conducted too late and not on initial production units, and a Risk Management Matrix that failed to identify all relevant risks. The FDA requires prompt corrective actions. The firm must notify the FDA within fifteen business days with specific steps taken, prevention plans, and supporting documentation. Failure to correct these serious quality system deficiencies may result in regulatory actions such as seizure, injunction, civil money penalties, and impact on federal contracts, premarket approvals, and Certificates to Foreign Governments.

Company
Fresenius Medical Care, North America
Inspection date
December 7, 2012
Product type
Devices
Office
Denver District Office
People
  • Latonya Mitchell (District Director)
  • Steven D. Marler
View Company Profile

ID · 13ad9f83-444c-496f-9ebe-aabaadfbd30b

Violation Codes5
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(f)21 CFR 820.30(g)

Full citation text and observation details available on the Dashboard.

Other documents for Fresenius Medical Care, North America
  • 483 - 2026-03-06March 6, 2026
  • 483 - 2021-05-21May 21, 2021
  • 483 - 2020-08-13August 13, 2020

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Fresenius Medical Care, North America
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.