WARNING LETTER
•GUANGDONG BIOLIGHT MEDITECH CO., LTD•August 23, 2012

GUANGDONG BIOLIGHT MEDITECH CO., LTD – FDA Warning Letter, Aug 23, 2012

Record Details

The FDA issued a Warning Letter to Guangdong Biolight Meditech Co., Ltd. following an inspection from August 20-23, 2012, at their Zhuhai, China facility, which manufactures physiological patient monitors and accessories. The inspection revealed that the firm's devices are adulterated under 21 U.S.C. § 351(h) because their manufacturing, packing, storage, or installation methods do not conform to the Quality System (QS) regulation, 21 CFR Part 820. Additionally, devices were found to be misbranded under 21 U.S.C. § 352(t)(2) and § 352(b).

Key QS violations include failures to: maintain complaint files and procedures (21 CFR 820.198(a)), maintain investigation records (21 CFR 820.198(e)), establish design validation procedures (21 CFR 820.30(g)), establish design verification procedures (21 CFR 820.30(f)), implement corrective and preventive action procedures (21 CFR 820.100(a)), control nonconforming product (21 CFR 820.90(a)), conduct formal design reviews (21 CFR 820.30(e)), and maintain complete device history records (21 CFR 820.184).

Misbranding violations include the failure to develop and implement written Medical Device Reporting (MDR) procedures (21 CFR 803.17) and the failure of FingerTrip Pulse Oximeter labels to contain the manufacturer's name and place of business (21 U.S.C. 352(b), 21 CFR 801.1).

The firm's September 11, 2012, response to the FDA 483 was deemed inadequate as it lacked documentation of corrections and corrective actions. The FDA requires a written response within fifteen business days detailing specific steps taken, including systemic corrective actions, documentation, and a timetable for completion. Failure to correct violations may impact premarket approval applications, Certificates to Foreign Governments, and federal contracts.

Company
GUANGDONG BIOLIGHT MEDITECH CO., LTD
Inspection date
August 23, 2012
Product type
Devices
Office
Center for Devices and Radiological Health
Person
  • Steven D. Silverman (President)
View Company Profile

ID · fd94dcbc-2a36-4923-bcad-66f722737965

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.198(a)21 CFR 820.198(e)21 CFR 820.30(g)21 CFR 820.30(f)21 CFR 820.100(a)21 CFR 820.90(a)21 CFR 820.30(e)

Full citation text and observation details available on the Dashboard.

Other documents for GUANGDONG BIOLIGHT MEDITECH CO., LTD
  • 483 - 2024-09-05September 5, 2024

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore GUANGDONG BIOLIGHT MEDITECH CO., LTD
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.