Attention www.globalkeysolutions.net is now www.keypedia.com, read the full press release here.
KeyPedia LogoKeyPedia Icon
  1. Home
  2. /
  3. Records
  4. /
  5. Warning Letter
  6. /
  7. H & H Associates
WARNING LETTER
•H & H Associates•January 10, 2012

H & H Associates – FDA Warning Letter, Jan 10, 2012

Record Details

On April 30, 2012, the FDA issued a Warning Letter to H & H Associates following an inspection from December 1, 2011, to January 10, 2012, at their Ordinary, Virginia facility. The inspection determined that the firm manufactures the H & H Emergency Cricothyrotomy Kit and Chest Tube Insertion Kit, classified as devices under the Federal Food, Drug, and Cosmetic Act.

The H & H Emergency Cricothyrotomy Kit was found to be adulterated because the firm lacked an approved premarket approval (PMA) application or an investigational device exemption. It was also misbranded as the firm failed to notify the agency of its intent to introduce the device into commercial distribution via a 510(k) submission. The FDA noted that the kit requires 510(k) clearance, with documentation for performance, biocompatibility, clinical, and usability validation.

Both devices were deemed adulterated due to non-conformity with current good manufacturing practice requirements of the Quality System (QS) regulation (21 CFR Part 820). Specific QS violations included: 1. Failure to establish and maintain design control procedures (21 CFR 820.30(a)), including design input, output, review, verification, validation, change, and transfer, and lack of a Design History File. 2. Failure to validate processes where results cannot be

Company
H & H Associates
Inspection date
January 10, 2012
Product type
Devices
Office
Baltimore District Office
Person
  • Evelyn Bonnin (District Director)
View Company Profile

ID · 9bd5e8e8-6718-4a6c-9921-4c30476f6c74

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(f)(1)(B)21 U.S.C. 360e(a)21 U.S.C. 360j(g)21 U.S.C. 352(o)21 U.S.C. 360(k)21 CFR 807.81(b)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore H & H Associates
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.