WARNING LETTER
•HeartWare Inc•January 24, 2014

HeartWare Inc – FDA Warning Letter, Jan 24, 2014

Record Details

On January 13-24, 2014, an FDA inspection of HeartWare, Incorporated's Miami Lakes, Florida facility, which manufactures the HeartWare Ventricular Assist Device (HVAD) system, revealed the device to be adulterated. The manufacturing, packing, storage, or installation methods, facilities, or controls did not conform to current good manufacturing practice (cGMP) requirements of the Quality System (QS) regulation (21 CFR Part 820).

Violations included: 1. **Failure to validate device design (21 CFR 820.30(g)):** The firm's Design Control System Procedure did not require design validation of device labeling (Patient Manual) for patient/caregiver understanding. The firm's responses were inadequate as they did not detail retrospective validation activities for the HVAD system. 2. **Failure to establish and maintain CAPA procedures (21 CFR 820.100(a)):** * The firm failed to verify or validate the effectiveness of corrective actions for HVAD controller failures (27 complaints, 2 deaths, 4 serious injuries) attributed to ESD. * CAPA procedures were not adequately implemented for premature battery failure issues (238 complaints, MDR events). * Rework of controller inventory with an updated ESD shield design and field actions were not completed after the updated design's release.

Company
HeartWare Inc
Inspection date
January 24, 2014
Product type
Devices
Office
Florida District Office
People
  • Elizabeth W. Ormond
  • Ramon Paz
View Company Profile

ID · c59b3582-262e-4838-9980-4933fd3a0d19

Violation Codes7
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(g)21 CFR 820.100(a)21 CFR 820.198(e)21 CFR 820.70(i)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore HeartWare Inc
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.