Intelligent Hospital Systems – FDA Warning Letter, Sep 22, 2011
On September 19-22, 2011, an FDA inspection of Intelligent Hospital Systems in Winnipeg, Canada, revealed that their RIVA pharmaceutical compounding device was adulterated under 21 U.S.C. § 351(h). The firm's manufacturing, packing, storage, or installation methods, facilities, or controls did not conform to the Quality System regulation (21 CFR Part 820).
The Warning Letter cited several violations, including: 1. **Failure to establish and maintain adequate procedures for corrective and preventive action (21 CFR 820.100(a))**: The Corrective and Preventive Action Procedure, QAP0600, lacked definitions for CAPA investigation, identifying actions to prevent recurrence, and verifying/validating effectiveness. The firm's response was inadequate, as the updated procedure still lacked details on identifying recurrence prevention actions, and no evidence of implementation or training was provided. 2. **Failure to establish and maintain adequate procedures for receiving, reviewing, and evaluating complaints (21 CFR 820.198(a))**: Procedure FDA0903, used by call center and technical support employees, did not define the timeframe or criteria for determining what constitutes a complaint, documentation requirements, or reference the firm's Complaint Procedure, QAP0500. The adequacy of the firm's response could not be determined due to
ID · 99e2f113-9b1f-4bab-902c-ff02c4ac67da
Full citation text and observation details available on the Dashboard.
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.