FDA WARNING_LETTER - Invacare Corporation - June 24, 2021
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The FDA issued a Warning Letter to Invacare Way following an inspection from May 25 to June 24, 2021, at their medical device operations in Elyria, OH. The inspection revealed that Invacare, a manufacturer of Class II Oxygen Concentrators, failed to conform to Quality System regulation (21 CFR Part 820), rendering their devices adulterated under section 501(h) of the Act.
Key violations include: 1. **Failure to investigate complaints (21 CFR § 820.198(c)):** Invacare did not investigate approximately 10% of Oxygen Concentrator complaints (2,589/25,777) between February 2020 and April 2021, particularly for devices past their end-of-life or component wear period, or when component failure was assumed due to wear. The firm's responses and corrective actions (CAPA CA-00067) were deemed inadequate as the updated procedure still doesn't require investigations regardless of injury/death. 2. **Inadequate complaint procedures (21 CFR § 820.198(a)):** The "Customer Feedback and Complaint Entry" procedure (WI-CP14-002-022) inadequately assigns severity scores for harms like hypoxia, leading to devices not being returned for investigation when
ID · 932f3372-9d8b-4357-9b99-e2e829194e1c