FDA WARNING_LETTER - Lac-Mac Limited - November 06, 2014
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On November 3-6, 2014, an FDA inspection of Lac Mac, Ltd. in London, Ontario, identified that their surgical gowns, sterile wraps, and surgical drapes are adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820). The firm's November 25, 2014, response to the FDA 483 was deemed inadequate.
Key violations include: 1. **Design Control (21 CFR 820.30(a)):** Failure to establish and maintain procedures ensuring design input/output, verification, validation, review, changes, and transfer are documented. DHFs lacked this information. The firm's revised procedures were insufficient without a retrospective DHF review. 2. **CAPA (21 CFR 820.100(a)):** Inadequate CAPA procedures lacking requirements for data analysis, investigation, action identification, verification/validation, implementation, information dissemination, management review submission, and documentation. Specific CAPA records (137, 138, 143, 144) lacked investigation results, validation, and implementation dates. CAPA 138, concerning barcode complaints, lacked documented verification/validation of corrective actions. The firm's revised CAPA
ID · 934728be-3a7f-4b4f-951e-77ea5eb3f444