Record Details

The FDA issued a Warning Letter to Med-Mizer, Inc. following a February-March 2014 inspection of their Batesville, Indiana facility, which manufactures AC powered, adjustable, and bariatric hospital beds. The inspection revealed the firm's devices are adulterated under 21 U.S.C. § 351(h) due to widespread non-conformity with the Quality System regulation (21 CFR Part 820).

Significant Quality System violations include failures in establishing and maintaining procedures for corrective and preventive actions, complaint handling, production process control, and comprehensive design control (inputs, outputs, reviews, verification, validation, and change control). Deficiencies also covered process validation, finished device acceptance, nonconforming product control, rework, purchasing controls, incoming product acceptance, Device History Records, and employee training.

The firm's Model SS Hospital Beds were misbranded for failing to report a field correction/removal (21 CFR Part 806.10), though a voluntary recall and report were later initiated. Misbranding also stemmed from lacking Medical Device Reporting (MDR) procedures (21 CFR 803.17) and failing to submit an MDR for a patient death within 30 days (21 CFR 803.50(a)(1)).

Med-Mizer's initial response was largely inadequate. The FDA requires prompt, systemic corrective actions, with a detailed written response within 15 business days, including documentation and a realistic timetable. Non-compliance may lead to severe regulatory actions, including seizure, injunction, civil money penalties, and adverse impacts on federal contracts and device approvals.

Company
Med-Mizer, Inc.
Inspection date
March 7, 2014
Product type
Devices
Office
Detroit District Office
Person
  • Paige E. Shelborne
View Company Profile

ID · 5c073484-385e-4160-90c4-8c04eae59634

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.100(a)21 CFR 820.100(a)(1)21 CFR 820.100(a)(3)21 CFR 820.100(a)(4)21 CFR 820.19821 CFR 820.198(a)(3)21 CFR 820.198(b)

Full citation text and observation details available on the Dashboard.

Other documents for Med-Mizer, Inc.
  • 483 - 2021-08-06August 6, 2021

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Med-Mizer, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.