Meridian Co., Ltd. – FDA Warning Letter, May 29, 2014
On August 19, 2014, the FDA issued a Warning Letter to Meridian Co., Ltd. following an inspection from May 26-29, 2014, which found their infrared lamps and plethysmographs to be adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820). The firm's June 12, 2014 response to the FDA 483 was deemed inadequate.
Key violations include: 1. **Complaint Handling (21 CFR 820.198(a)):** Failure to classify product malfunctions as complaints, evaluate for MDR reportability, and provide documentation of revised procedures and training. 2. **Corrective and Preventive Action (CAPA) (21 CFR 820.100(a)):** CAPA procedure lacked verification/validation requirements, and nonconformance reports were not analyzed for root causes. Response lacked implementation documentation. 3. **Purchasing Controls (21 CFR 820.50):** Exempted components from supplier qualification without specification, lacked contracts with major suppliers, and failed to maintain records of acceptable suppliers. No corrective action description provided. 4. **Equipment Calibration (21 CFR 820.72(
ID · 91df6d84-b9cf-4ef1-a151-ea459564df65
Full citation text and observation details available on the Dashboard.
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.