WARNING LETTER
•Millar Instruments, Inc.•February 2, 2012

Millar Instruments, Inc. – FDA Warning Letter, Feb 2, 2012

Record Details

On May 7, 2012, the FDA issued a Warning Letter to Millar Instruments, Inc. following an inspection from January 24 to February 2, 2012. The inspection revealed that the firm's reusable and disposable pressure transducer catheters, including urological, cardiovascular, and intracranial models, are adulterated under 21 U.S.C. § 351(h) because their manufacturing, packing, storage, or installation methods do not conform to the Quality System regulation (21 CFR Part 820).

Key violations include: 1. **Failure to validate processes (21 CFR 820.75(a)):** Processes for preparing catheter surfaces for adherence were not validated with a high degree of assurance, lacking protocols, acceptance criteria, worst-case testing, and verification of intended use. The firm's response was deemed inadequate as it did not provide referenced documents or sufficient supporting documentation for proposed corrective actions. 2. **Failure to establish and maintain procedures for finished device acceptance (21 CFR 820.80(d)):** The firm used non-validated in-house test methods for critical in-process and final device testing. Validation of test equipment and methods was inadequate, lacking protocols, acceptance criteria, and worst-case testing. The firm's response was inadequate, failing to provide supporting documentation for test method validity or proposed corrective actions. 3

Company
Millar Instruments, Inc.
Inspection date
February 2, 2012
Product type
Devices
Office
Department of Health and Human Services
Person
  • Reynaldo Rodriguez (District Director)
View Company Profile

ID · 27cf677c-0358-413e-87e6-db47242b10d4

Violation Codes8
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 820.75(a)21 CFR 820.80(d)21 CFR 820.18421 CFR 820.70(c)21 CFR 820.250(b)21 CFR 820.22

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Millar Instruments, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.