WARNING LETTER
•MTI Precision Products LLC.•April 30, 2014

MTI Precision Products LLC. – FDA Warning Letter, Apr 30, 2014

Record Details

An FDA inspection of MTI Precision Products LLC (MTI Dental Products) from April 17-30, 2014, revealed that their manufactured dental devices, including Lynx handpieces and Sonic Scalers, are adulterated. The firm's manufacturing methods, facilities, or controls do not conform with the Quality System (QS) regulation (21 CFR Part 820). The FDA identified ten significant QS regulation violations, including failures to establish and maintain adequate procedures for corrective and preventive action, design changes, complaint handling (both general and for Medical Device Reporting), purchasing controls, nonconforming product control, rework, device history records, quality audits, and personnel training. Additionally, the Lynx TM20 TorqueMaster low speed dental handpiece was found to be misbranded due to the firm's failure to report a device correction or removal to the FDA, as required by 21 CFR 806.10(a), following a recall initiated due to a sharp edge defect. MTI's response to the Form FDA 483 observations was deemed inadequate for all violations. The FDA requires prompt corrective action within fifteen working days, including documentation and a timetable for future corrections. Failure to comply may result in regulatory actions such as seizure, injunction, civil money penalties, impact on federal contracts, and denial of premarket approvals or Certificates to Foreign Governments.

Company
MTI Precision Products LLC.
Inspection date
April 30, 2014
Product type
Devices
Office
New Jersey District Office
Person
  • Diana Amador-Toro (District Director)
View Company Profile

ID · 744063fe-d663-4d0b-921f-dcbd11d4e543

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.100(a)21 CFR 820.30(i)21 CFR 820.198(a)(3)21 CFR 820.198(a)21 CFR 820.5021 CFR 820.90(a)21 CFR 820.90(b)(2)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore MTI Precision Products LLC.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.