Navilyst Medical, Inc. – FDA Warning Letter, May 28, 2014

Record Details

This FDA Warning Letter, NYK-2015-5, issued to Navilyst Medical Inc., an AngioDynamics Company, details violations observed during inspections from April-May 2014 (Marlborough, MA) and February-March 2014 (Glens Falls, NY). The company is a specification developer and manufacturer of Vaxcel, Xcela, and Biostable (BioFlo) Peripherally Inserted Central Catheters (PICC) medical devices. These devices are deemed adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation, 21 CFR Part 820.

Key violations at the Marlborough, MA facility include: 1. **Failure to establish and maintain procedures for verifying device design (21 CFR 820.30(f))**: Issues with seal width specifications for packaged kits, inadequate packaging design verification for protection and sterile barrier maintenance (e.g., cracked trays, holes), and insufficient accelerated/real-time aging testing procedures (e.g., out-of-tolerance humidity, product release before real-time study completion, retesting failures without documentation). 2. **Failure to establish and maintain design requirements (21 CFR 820.30(c))**: Design verification for Xcela PASV and BioFlo w/

Company
Navilyst Medical, Inc.
Inspection date
May 28, 2014
Product type
Devices
Office
New York District Office
Person
  • Ronald M. Pace (Program Director)
View Company Profile

ID · a244ce4c-5bda-41df-8719-79eb0c64385a

Violation Codes9
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(f)21 CFR 820.30(c)21 CFR 820.10021 CFR 820.2221 CFR 820.75(a)21 CFR 820.90(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Navilyst Medical, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.