Neitz Instruments Company Ltd. – FDA Warning Letter, May 19, 2023

Record Details

On August 25, 2023, the FDA issued a Warning Letter to a firm in Ichibancho, Japan, following an inspection from May 15-19, 2023. The firm manufactures the Neitz Halogen Ophthalmoscope BXa, 13A, which are medical devices. The inspection revealed the devices are misbranded under section 502(t)(2) of the Act, due to the firm's failure to furnish required material or information under section 519 of the Act and 21 CFR Part 803 - Medical Device Reporting (MDR).

Significant violations include the failure to develop, maintain, and implement written MDR procedures as required by 21 CFR 803.17. Specifically, the firm's procedures lacked: 1. Internal systems for timely identification, communication, and evaluation of events subject to MDR, including definitions for reportable events (e.g., "serious injury," "malfunction") and instructions for timely evaluation. 2. A standardized review process for determining when an event meets reporting criteria, including instructions for investigating reportable events and identifying who makes reporting decisions. 3. Internal systems for timely transmission of complete MDRs, such as instructions for completing FDA Form 3500A, circumstances for submitting various report types (30-day, 5-day, follow-up), and a process for

Company
Neitz Instruments Company Ltd.
Inspection date
May 19, 2023
Product type
Devices
Office
Center for Devices and Radiological Health
People
  • Malvina B. Eydelman (Office Director)
  • Takahiro Haruyama
View Company Profile

ID · dc39e87f-595d-45e2-a1e2-9947acdd51f1

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 80321 CFR 803.1721 CFR 803.17(a)(1)21 CFR 803.50(a)21 CFR 803.17(a)(2)21 CFR 803.17(a)(3)21 CFR 803.12(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Neitz Instruments Company Ltd.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.