WARNING LETTER
•Ophthalmic Technologies, Inc.•August 2, 2007

FDA WARNING_LETTER - Ophthalmic Technologies, Inc. - August 02, 2007

Record Details

An FDA inspection of Ophthalmic Technologies, Inc. (OTI) in Toronto, Canada, from July 30 to August 2, 2007, found that the firm's OTI Scan 1000, OTI Scan 2000, and OTI OCT/SLO imaging systems are adulterated. This is due to non-conformity with Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation (21 CFR Part 820) and a failure to establish Medical Device Reporting (MDR) procedures (21 CFR 803.17).

Key QS violations include failures in establishing and maintaining adequate procedures for corrective and preventative actions (CAPA), complaint handling, incoming product acceptance, quality audits, management review, and personnel training (21 CFR 820.100, 820.198, 820.80(b), 820.22, 820.20(c), 820.25(b)). Additionally, OTI failed to develop and implement written MDR procedures for timely event identification, evaluation, and transmission (21 CFR 803.17(a)(1)).

The firm's August 28, 2007, response was deemed inadequate. Consequently, OTI's devices are subject to refusal of admission under an Import Alert. Federal agencies will be informed, impacting contract awards. Premarket approvals for Class III devices and Certificates to Foreign Governments will be withheld until all violations are corrected. OTI must provide a written response within 15 working days detailing corrective actions and a timeline.

Company
Ophthalmic Technologies, Inc.
Inspection date
August 2, 2007
Product type
Devices
Office
Center for Devices and Radiological Health
People
  • Timothy A. Ulatowski (Director)
  • Jon Schnaars
View Company Profile

ID · 853627fd-c155-4521-a0b8-5ac562ee2697

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.100(a)(4)21 CFR 820.100(a)(3)21 CFR 820.100(a)(5)21 CFR 820.19821 CFR 80321 CFR 820.198(b)21 CFR 820.198(c)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Ophthalmic Technologies, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.