•Orthotics & Prosthetics One Inc•December 1, 2011

Orthotics & Prosthetics One Inc – FDA Warning Letter, Dec 1, 2011

Record Details

On November 29-December 1, 2011, an FDA inspection of Orthotics & Prosthetics One Inc. in Waterloo, IA, identified significant violations of the Quality System (QS) regulation for medical devices. The firm manufactures lower limb prosthetics, classified as Class II medical devices (21 CFR 890.3500), which are exempt from premarket notification but subject to current good manufacturing practice (CGMP) requirements (21 CFR Part 820).

The inspection revealed that the devices are adulterated under section 501(h) of the Act due to non-conformity with CGMP. Specific violations include: 1. Failure to establish a quality plan (21 CFR 820.20(d)). 2. Failure to establish quality system procedures and instructions (21 CFR 820.20(e)). 3. Failure to establish a quality policy and quality objectives (21 CFR 820.20(e)). 4. Failure to perform quality audits (21 CFR 820.22). 5. Failure to establish procedures for purchased/received product and service conformity (21 CFR 820.50). 6. Failure to establish a device master record (21 CFR 820.181). 7. Failure to establish design plans (21 CFR 82

Company
Orthotics & Prosthetics One Inc
Inspection date
December 1, 2011
Product type
Devices
Office
Kansas City District Office
Person
  • John W. Thorsky (District Director)
View Company Profile

ID · 5f848910-cc1b-48f6-ad94-f2cf60942a3e

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 890.350021 CFR 82021 CFR 820.20(d)21 CFR 820.20(e)21 CFR 820.2221 CFR 820.5021 CFR 820.18121 CFR 820.30(b)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Orthotics & Prosthetics One Inc
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.