•Philips Medical Systems•August 9, 2010

Philips Medical Systems – FDA Warning Letter, Aug 9, 2010

Record Details

This FDA Warning Letter, dated March 28, 2011, is an amendment to a previous letter from March 22, 2011, correcting typographical errors and extending the response deadline. It addresses violations observed during an inspection from April 5, 2010, to August 9, 2010, at a Seattle, Washington firm manufacturing HeartStart, FRx, FR2+, and HS1 lines of Automated External Defibrillators (AEDs).

The inspection revealed that the devices are adulterated under 21 U.S.C. § 351(h) due to non-conformity with Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation (21 CFR Part 820). The firm's responses from August 30, 2010, and October 21, 2010, were deemed inadequate for most violations.

Key violations include: 1. **Failure to adequately verify or validate CAPA effectiveness (21 CFR 820.100(a)(4))**: Evidenced by 87 complaints, including one death, related to detached/missing magnets on HS1 AED Pads Cartridges, despite multiple CAPAs and design changes lacking effectiveness validation. 2. **Failure to adequately implement CAPA changes (21 CFR 820.1

Company
Philips Medical Systems
Inspection date
August 9, 2010
Product type
Devices
Office
Seattle District Office
People
  • Charles M. Breen (District Director)
  • Daniel Donnelly
View Company Profile

ID · 9f285c9d-3a88-401f-b538-118158f4c6dc

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.100(a)(4)21 CFR 820.100(a)(5)21 CFR 820.30(g)21 CFR 820.70(i)21 CFR 820.100(b)21 CFR 820.198(a)21 CFR 820.198(e)(3)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Philips Medical Systems
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.