Probed Medical Technologies, Inc. – FDA Warning Letter, Sep 15, 2011

Record Details

On September 12-15, 2011, an FDA inspection of Probed Medical Technologies, Inc. in Abbotsford, Canada, revealed that their device, the Freedom Bed, was adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation 21 CFR Part 820.

The inspection identified four key violations: 1. **Failure to establish and maintain adequate procedures for design changes (21 CFR 820.30(i)):** The firm failed to document validation/verification of design changes, such as a bolt size change for a "clunk" noise and a design change in October 2008, to ensure effectiveness and no adverse effects on the finished product. The firm's response was inadequate as it lacked a modified procedure and retrospective review documentation. 2. **Failure to establish and maintain adequate procedures for finished device acceptance (21 CFR 820.80(d)):** Written procedures for finished product testing, specifically MP018, did not cover hand control, side rail, and battery charger testing. The firm's response was inadequate, lacking evidence of revised procedures, training, or comprehensive testing documentation. 3. **Failure to establish and maintain adequate requirements for suppliers (21 CFR 820.50(a)(2)):

Company
Probed Medical Technologies, Inc.
Inspection date
September 15, 2011
Product type
Devices
Office
Center for Devices and Radiological Health
Person
  • Steven D. Silverman (President)
View Company Profile

ID · 886b72bc-7f64-4d21-8422-12af9b55e981

Violation Codes8
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(i)21 CFR 820.80(d)21 CFR 820.50(a)(2)21 CFR 820.2221 U.S.C. 381(a)

Full citation text and observation details available on the Dashboard.

Other documents for Probed Medical Technologies, Inc.
  • 483 - 2022-08-11August 11, 2022

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Probed Medical Technologies, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.