Record Details

On December 2, 2011, the FDA inspected Prosec Protection Systems, Inc. in Lakewood, New Jersey, and found that their My Child Identification Kit, a medical device, was adulterated. The inspection revealed that the company's manufacturing, packing, storage, or installation methods, facilities, or controls did not conform to the Quality System regulation (21 CFR Part 820).

Violations included: 1. **Failure to establish finished device acceptance procedures (21 CFR 820.80(d)):** No pass/fail results were recorded for sterilization records, and sterilization cycle reports were unsigned and contained mismatched parameters. 2. **Failure to validate processes (21 CFR 820.75(a)):** Sterilization validation protocols lacked critical parameters (time, temperature, pressure, plasma, gas concentration), validation runs were not consecutive, and a failed interim run was not investigated. Protocols were also not signed by management prior to validation. 3. **Failure to establish procedures for monitoring and control of process parameters (21 CFR 820.75(b)):** Sterility testing procedures did not include positive growth controls, and biological indicators were not placed horizontally as per manufacturer instructions. 4. **Failure to establish corrective and preventive action (CAPA) procedures (21 CFR 820.100(a)):** No written SOP for documenting CAPAs in the software system

Company
Prosec Protection Systems, Inc.
Inspection date
December 22, 2011
Product type
Devices
Office
New Jersey District Office
Person
  • Diana Amador-Toro (District Director)
View Company Profile

ID · 351642c6-72be-4056-b822-c4532a60f26d

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.80(d)21 CFR 820.75(a)21 CFR 820.75(b)21 CFR 820.100(a)21 CFR 820.90(a)21 CFR 820.20(b)21 CFR 820.70(i)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Prosec Protection Systems, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.