Racer Technology Pte Ltd – FDA Warning Letter, Jun 6, 2013
On June 3-6, 2013, the FDA inspected Racer Technology Pte. Ltd. in Singapore, finding its cochlear implant accessories (earhooks, battery chargers) adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820). The firm's June 19, 2013, response to the FDA 483 was deemed inadequate for multiple violations.
Key violations include: 1. **Failure to establish and maintain procedures for changes (21 CFR 820.70(b))**: The firm lacked qualification records for a new (b)(4) and documentation for its implementation. The response did not address verification/validation and documentation deficiencies. 2. **Failure to establish and maintain CAPA procedures (21 CFR 820.100(a))**: The procedure RACER-QP16 lacked requirements for analyzing quality data, using statistical methodology, and verifying/validating CAPAs. The updated procedure still omitted statistical methodology and verification/validation requirements. 3. **Failure to document CAPA activities (21 CFR 820.100(b))**: RACER-QP16 did not ensure documentation of all CAPA activities (analysis, investigation, actions, verification/validation, change control,
ID · 5a05222a-d8ba-4674-8882-efac1437a5b3
Full citation text and observation details available on the Dashboard.
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.