On February 20, 2015, the FDA issued a Warning Letter to Rodo Medical, Inc. following an inspection from July 29 to August 1, 2014. The inspection assessed compliance with federal regulations for a clinical feasibility study of the Rodo Dental Abutment System, titled "(b)(4)," conducted prior to an Investigational Device Exemption (IDE) submission.

The FDA identified several violations of 21 CFR Part 812 - Investigational Device Exemptions:

1. **Failure to obtain FDA approval for a significant risk device study (21 CFR 812.40, 812.42, 812.20(a)(1)):** Rodo Medical treated nine subjects with the investigational device between August 21, 2011, and March 1, 2013, without an approved IDE. The FDA had previously informed Rodo Medical in August 2011 that the device was considered significant risk due to potential bacterial ingress and required an IDE. Rodo Medical's response was deemed inadequate for not demonstrating understanding of device risk determination and proper disclosure to the IRB.

2. **Failure to properly monitor an investigation (21 CFR 812.2(b)(1)(iv)):** Rodo Medical did not ensure the clinical investigator complied with the signed agreement regarding UCLA Medical IRB

Company
Rodo Medical, Inc.
Inspection date
August 1, 2014
Product type
Devices
Office
Center for Devices and Radiological Health
Person
  • Jan B. Welch
View Company Profile

ID · a9a84d10-36c9-4272-a6ba-da3f926203e0

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 360j(g)21 CFR 812.4021 CFR 812.4221 CFR 812.20(a)(1)21 CFR 812.3(m)(4)21 CFR 812.2(b)21 CFR 812.2(b)(1)(iv)21 CFR 812.4621 CFR 812.2(b)(1)(v)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Rodo Medical, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.