FDA WARNING_LETTER - Royal Philips - October 26, 2023
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The FDA issued a Warning Letter to Philips Healthcare (Suzhou) Co., Ltd. following an inspection from October 23-26, 2023, at their Suzhou, China facility, which manufactures Computed Tomography (CT) and ultrasonography systems. The inspection found these devices adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and misbranded under 21 U.S.C. § 352(t)(2) for failing to report corrections and removals (21 CFR Part 806).
Key violations include: 1. **Failure to adequately control suppliers (21 CFR 820.50(a)(2)):** The firm failed to ensure validation of the injection molding process by a supplier for the Philips Incisive CT Patient Interface Monitor (PIM) data cable. This led to 64 complaints, a cable redesign, and a field correction due to ECG gating signal loss/error, potentially causing rescans and diagnostic delays. The firm's responses were deemed inadequate as the retrospective review was limited in scope (July 2021-August 2022) and lacked documentation of employee training on revised protocols. 2. **Failure to submit written reports of corrections or removals (21 CFR 80
ID · 790d15d9-0a92-43b9-833e-ded5cd36db00
Full citation text and observation details available on the Dashboard.