FDA WARNING_LETTER - Sea-Long Medical Systems, LLC - September 08, 2022
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The FDA issued a Warning Letter to Sea-Long Medical Systems, LLC, following an inspection from August 22 to September 8, 2022. The firm manufactures Sea-Long Medical Treatment Hoods (latex and non-latex, single and multi-patient neck seals with hoods). These devices are deemed adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and misbranded under 21 U.S.C. § 352(o), § 352(c), and § 352(t)(2).
Key violations include: 1. **Complaint Handling (21 CFR 820.198(a)):** Failure to establish and maintain adequate complaint procedures, lacking documentation of investigation details, root cause analysis, and corrective/preventive actions for five complaint records. Missing complaint logs and product/serial IDs. 2. **Process Validation (21 CFR 820.75(a)):** Failure to validate processes where results cannot be fully verified by inspection and test. 3. **Design Change (21 CFR 820.30(i)):** Absence of established procedures for design changes for Treatment Hoods. 4. **Corrective and Preventive Action (CAPA) (21 CFR
ID · 4da7446c-ff6d-488e-85f0-8ae9f13f50a9
Full citation text and observation details available on the Dashboard.