•Siemens Healthcare Diagnostics•December 29, 2011

Siemens Healthcare Diagnostics – FDA Warning Letter, Dec 29, 2011

Record Details

The FDA issued a Warning Letter to Siemens Healthcare Diagnostics following an inspection from October 24, 2011, to December 29, 2011, at their Tarrytown, New York facility, which manufactures the ADVIA Centaur iPTH immunoassay. The FDA determined these devices are adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).

Key violations include: 1. **Design Requirements (21 CFR 820.30(c)):** Failure to establish and maintain adequate procedures for design requirements. For example, a design project lacked documentation of a formal design review at the R1 Phase, which was "N/A (Waived)." 2. **Design Review (21 CFR 820.30(e)):** Failure to ensure formal documented design reviews are planned and conducted at appropriate stages. Procedures allowed combining or eliminating design reviews, not ensuring proper review stages. 3. **Design Changes (21 CFR 820.30(i)):** Failure to establish adequate procedures for identification, documentation, validation, review, and approval of design changes. Regulatory assessments for pre-market submissions were performed before sufficient performance data was collected, and assessments were not conducted prior to implementing changes. 4. **Design

Company
Siemens Healthcare Diagnostics
Inspection date
December 29, 2011
Product type
Devices
Office
New York District Office
Person
  • Ronald M. Pace (Program Director)
View Company Profile

ID · bcc61d28-49bd-4b52-b5de-b4091c877da8

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 820.30(c)21 CFR 820.30(e)21 CFR 820.30(i)21 CFR 820.30(b)21 CFR 820.30(f)21 CFR 820.198(b)21 U.S.C. 351(f)(1)(B)21 U.S.C. 360e(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Siemens Healthcare Diagnostics
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.