Attention www.globalkeysolutions.net is now www.keypedia.com, read the full press release here.
KeyPedia LogoKeyPedia Icon
  1. Home
  2. /
  3. Records
  4. /
  5. Warning Letter
  6. /
  7. SpecLinc, Inc.
WARNING LETTER
•SpecLinc, Inc.•June 2, 2011

SpecLinc, Inc. – FDA Warning Letter, Jun 2, 2011

Record Details

During a May 31-June 2, 2011, inspection, the FDA found SpecLinc, Inc., a manufacturer of OTREC® otoscope specula, in violation of Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation (21 CFR Part 820). The firm's products are considered adulterated under Section 501(h) of the Act.

Key violations include: 1. **Complaint Handling (21 CFR 820.198(a)):** Failure to establish and maintain procedures for receiving, reviewing, and evaluating complaints, including determining if Medical Device Reporting (MDR) is required. No complaint procedure or files exist. An example cited a May 24, 2011, complaint about "damaged otoscope speculums" with "damaged tips" that "cannot be safely used," which was only documented by a note about "flash on tip of specula" and reimbursement. 2. **Corrective and Preventive Action (CAPA) (21 CFR 820.100(a)):** Failure to establish CAPA procedures, including analyzing data sources, investigating nonconformities, identifying corrective actions, verifying effectiveness, implementing changes, disseminating information, and submitting for management review. 3. **Production Process Controls (21 CFR 820.70(a)):** Failure

Company
SpecLinc, Inc.
Inspection date
June 2, 2011
Product type
Devices
Office
New Orleans District Office
Person
  • Monica R. Maxwell (Program Division Director)
View Company Profile

ID · ecc2272f-c160-4966-8c3a-e27773e48a71

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.198(a)21 CFR 80321 CFR 820.198(a)(3)21 CFR 820.198(b)21 CFR 820.100(a)21 CFR 820.70(a)21 CFR 820.181

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore SpecLinc, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.