WARNING LETTER
•Steris Corporation•February 9, 2011

Steris Corporation – FDA Warning Letter, Feb 9, 2011

Record Details

The FDA issued a Warning Letter to STERIS Corporation regarding its Verify SixCess Class 6 Challenge Packs and Chemical Indicators. The FDA reviewed promotional materials and the company's website, finding that STERIS was making claims for these devices that were not cleared during their 510(k) submissions.

Specifically, the cleared intended uses for these chemical indicators and challenge packs involve monitoring specific steam sterilization cycle parameters. However, STERIS's promotional materials, including package inserts, brochures, and flyers, contained objectionable claims. These claims implied that the devices could be used for "immediate release of all loads," "eliminate 'blind' release," "measure Sterility Success," and were "cleared for the release of all loads & all items in the load, regardless of content," including loads with implants.

The FDA stated that these claims represent a major modification to the devices' Intended Use and Indications for Use, requiring a new 510(k) submission under 21 CFR Part 807.81(a)(3)(ii). The letter emphasized that chemical indicators do not replace biological indicators, which monitor microbicidal function, and that STERIS's representation of FDA's position on Class 6 indicators for implant loads was misleading.

These violations cause the devices to be misbranded under section 502(o) of the Federal Food, Drug and Cosmetic Act. STERIS was required to notify the FDA within 15

Company
Steris Corporation
Inspection date
February 9, 2011
Product type
Devices
Office
Center for Devices and Radiological Health
Person
  • Steven D. Silverman (President)
View Company Profile

ID · bcef0818-73af-4845-ad79-60a5fed30b2c

Violation Codes2
21 CFR 807.81(a)(3)(ii)21 U.S.C. 502(o)

Full citation text and observation details available on the Dashboard.

Other documents for Steris Corporation
  • WARNING_LETTER - 2008-06-17June 17, 2008
  • WARNING_LETTER

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Steris Corporation
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.