On March 11, 2010, the FDA issued a Warning Letter to Talisman Ltd. following an inspection from November 2-12, 2009. The inspection revealed that the firm's Automated Blood Grouping and Antibody Test Systems, classified as medical devices, were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) Regulation (21 C.F.R. Part 820).

The FDA found Talisman Ltd.'s December 4, 2009, and March 5, 2010, responses inadequate as they did not address all observations from the FDA-483. Violations included: 1. Failure to justify not reporting corrective/removal actions to FDA for distributed software patches (21 C.F.R. § 806.20(b)(4)). 2. Failure to establish procedures for conducting quality audits, specifically lacking validation evidence for manufacturing software (21 C.F.R. § 820.70(i)). 3. Failure to have complete and implemented design control procedures, with the SOP not fully addressing design plan and input requirements (21 C.F.R. § 820.30(a)).

Company
Talisman Limited
Inspection date
November 12, 2009
Office
Baltimore District Office
Person
  • Evelyn Bonnin (District Director)
View Company Profile

ID · 7ca57501-393f-48bd-8470-1952b4bbc2c2

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 806.20(b)(4)21 CFR 820.70(i)21 CFR 820.30(a)21 CFR 820.2221 CFR 820.198(a)21 CFR 820.25(b)21 CFR 820.100

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Talisman Limited
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.