TALON, an S & S Technology Company – FDA Warning Letter, Mar 28, 2019
On August 2, 2019, the FDA issued a Warning Letter to TALON, an S&S Technology Company, following an inspection from March 11-28, 2019. The inspection revealed that TALON, a manufacturer of Class I & II medical devices, had adulterated devices due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations include: 1. **Failure to document CAPA activities and results (21 CFR 820.100(b))**: CAPAs were closed without verifying corrective actions, e.g., for internal audit effectiveness, risk analysis completion, and updated job descriptions. 2. **Failure to establish design change procedures (21 CFR 820.30(i))**: Design changes were implemented without documented verification or validation, specifically for the Class II VAC-FIX System. 3. **Failure to control nonconforming product (21 CFR 820.90(a))**: Procedures lacked requirements for evaluating nonconformances and initiating investigations, a repeat observation from 2011. 4. **Failure to maintain device history records (21 CFR 820.184)**: No batch records or lot/batch numbers for sub-components, and approximately 1,000 Class II VAC-FIX Systems lacked UDI numbers or production
ID · 9249a859-adcb-4626-af9a-e7aade10b26b
Full citation text and observation details available on the Dashboard.
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.