WARNING LETTER
•Thoratec Corporation•September 7, 2011

Thoratec Corporation – FDA Warning Letter, Sep 7, 2011

Record Details

On January 3, 2012, the FDA issued a Warning Letter to Thoratec Corporation following an inspection from August 22 to September 7, 2011, at their Pleasanton, California facility. The inspection revealed that Thoratec's Ventricular Assist Device Systems (Thoratec, HeartMate XVE, and HeartMate II) were misbranded under Section 502(t)(2) of the Federal Food, Drug, and Cosmetic Act due to failures in Medical Device Reporting (MDR) as required by 21 CFR Part 803.

Significant violations included: 1. Failure to report a death within 90 calendar days (Complaint 201008-0145, aware April 30, 2010, MDR received August 17, 2010). 2. Failure to report a device malfunction likely to cause death or serious injury if it recurred, within 90 calendar days (Complaint 201011-0160, aware November 11, 2010, MDR received March 21, 2011).

Thoratec's September 23, 2011, response regarding MDRs was deemed inadequate as it did not sufficiently address steps to prevent future late reporting.

Additionally, the FDA noted quality system deficiencies

Company
Thoratec Corporation
Inspection date
September 7, 2011
Product type
Devices
Office
San Francisco District Office
Person
  • Barbara J. Cassens (Director )
View Company Profile

ID · 53c9b2f7-d2f3-4226-a2e8-d8ebd1019264

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 80321 CFR 803.50(a)(1)21 CFR 803.19(e)21 CFR 803.50(a)(2)21 CFR 82021 CFR 820.75(a)21 CFR 820.70(g)(1)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Thoratec Corporation
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.