Record Details

An FDA inspection of Trokamed GmbH in Geisingen, Germany, from June 24-27, 2013, revealed that the firm's manufactured devices, including Intraluminal Artery Strippers and Morcellators, are adulterated. This is due to non-conformity with the Quality System (QS) regulation (21 CFR Part 820). The firm's July 17, 2013, response to the FDA 483 was largely deemed inadequate. Key violations include: failure to establish and maintain procedures for process validation (21 CFR 820.75(a)), with several manufacturing processes remaining unvalidated; failure to validate critical computer software for its intended use (21 CFR 820.70(i)); failure to establish and maintain procedures for the identification, documentation, validation, or verification, review, and approval of design changes before implementation (21 CFR 820.30(i)); failure to establish and maintain procedures to control all required documents (21 CFR 820.40), though the firm's response for this point appeared adequate; and failure to establish and maintain procedures for quality audits conducted by individuals without direct responsibility for audited matters (21 CFR 820.22). The FDA requires Trokamed to provide a written response within fifteen business days detailing specific corrective actions, prevention plans, and a timetable for completion, with supporting documentation. Failure to correct these violations may impact federal contracts, premarket approval for Class III devices, and Certificates to Foreign Governments. The letter emphasizes that these issues may indicate systemic problems in the firm's manufacturing and quality management systems.

Company
Trokamed GmbH
Inspection date
June 27, 2013
Product type
Devices
Office
Center for Devices and Radiological Health
Person
  • Steven D. Silverman (President)
View Company Profile

ID · 71fae3ea-8d53-45df-b619-5b1d98e05bbe

Violation Codes8
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.75(a)21 CFR 820.70(i)21 CFR 820.30(i)21 CFR 820.4021 CFR 820.22

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Trokamed GmbH
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.