Record Details

On February 18, 2011, the FDA issued a Warning Letter to Tytan Medical, Corp. following an inspection from October 25-28, 2010, at their Shanghai, China facility. The inspection revealed that the firm's Class I stethoscopes and Class II manual sphygmomanometers are adulterated under section 501(h) of the Federal Food, Drug, and Cosmetic Act, as their manufacturing, packing, storage, or installation methods do not conform to the Quality System (QS) regulation (21 CFR Part 820).

Key violations include: 1. **Failure to establish and maintain design control procedures (21 CFR 820.30(a))**: No design control process was established for manual sphygmomanometers manufactured since 2004, and existing procedures from May 2006 were not implemented. Documentation for design plans, inputs, outputs, reviews, verification/validation, transfer, and history files was absent. 2. **Failure to adequately establish and maintain corrective and preventative action (CAPA) procedures (21 CFR 820.100(a))**: Procedure P-40008 did not specify manufacturing acceptance activity data for analysis. Data from critical sphygmomanometer gage accuracy tests (first and second (b)(4) point checks), defined by the A

Company
Tytan Medical, Corp.
Inspection date
October 28, 2010
Product type
Devices
Office
Department of Health and Human Services
Person
  • Steven D. Silverman (President)
View Company Profile

ID · 91d46ed8-1adf-44d4-b88f-4da7d8a51bb6

Violation Codes8
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(a)21 CFR 820.100(a)21 CFR 820.198(a)21 CFR 80321 U.S.C. 381(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Tytan Medical, Corp.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.