Uscom Europe, located in Budapest, Hungary, was inspected by the FDA from January 27 to January 30, 2025. The inspection revealed significant regulatory violations concerning the company"s spirometer devices, particularly the SpiroSonic AIR. This device is considered adulterated under section 501(f)(1)(B) of the Federal Food, Drug, and Cosmetic Act because it lacks an approved premarket approval (PMA) application or an investigational device exemption. Additionally, it is misbranded under section 502(o) for failing to notify the FDA of its commercial distribution intent, as required under section 510(k).

The SpiroSonic AIR"s design and intended use changes, including multi-patient use flow tubes, Bluetooth technology, and home setting applications, significantly affect its safety and effectiveness compared to the previously cleared SpiroThor device. Despite submitting a premarket notification (K220982) in April 2022, Uscom Europe did not adequately respond to the FDA"s request for additional information, leading to the withdrawal of their submission in February 2023. Consequently, the SpiroSonic AIR lacks 510(k) clearance and cannot be legally marketed.

The inspection also identified Quality System Regulation (QSR) violations under section 501(h), specifically the failure to establish and maintain adequate design validation procedures. The SpiroSonic Mobile Software V1.16, intended for use with SpiroSonic AIR, was not properly validated. The FDA requires Uscom Europe to cease activities that result in device misbranding or adulteration and to address the noted violations to comply with regulatory standards.

Company
Uscom Kft
Inspection date
January 30, 2025
Product type
Devices
Office
Office of Product Evaluation and Quality
Person
  • Kesia Alexander
View Company Profile

ID · 998ca010-17e9-4f76-9e75-d7f25d615833

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(f)(1)(B)21 U.S.C. 360e(a)21 U.S.C. 360j(g)21 U.S.C. 352(o)21 U.S.C. 360(k)21 CFR 807.81(b)21 U.S.C. 351(h)21 CFR 820.30(g)21 CFR 820.100(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Uscom Kft
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.