WARNING LETTER
•Visgeneer, Inc.•February 27, 2025

Visgeneer, Inc. – FDA Warning Letter, Feb 27, 2025

Record Details

The FDA issued a warning letter to Dr. Ken-Shwo Dai of the firm located at No. 335 Zhong Hua Rd., Sec. 6, Xiangshan District, Hsinchu City, Taiwan, following an inspection conducted from February 24 to February 27, 2025. The inspection revealed significant violations concerning the manufacturing processes of the eBchek Blood Glucose Monitoring System and the eBuricacid/Uritouch Blood Uric Acid Monitoring System. These products are classified as medical devices under section 201(h) of the Federal Food, Drug, and Cosmetic Act.

The primary issue identified was the failure to establish and maintain proper procedures for validating device design, as mandated by 21 CFR 820.30(g). Specifically, the firm lacked documentation for parts or software versions used in manufacturing, as well as production records for the units tested. The response from Dr. Dai, dated March 13, 2025, attempted to address these issues by referencing test reports and studies conducted on mass production units. However, the response was deemed inadequate due to missing information and inconsistencies in software version documentation.

The FDA highlighted discrepancies between the firm’s new analytical studies and the performance specifications cleared in the original 510(k) submission (k122181). These discrepancies included differences in precision and accuracy results, as well as the use of a different comparator method (YSI 2500 instead of YSI 2300) without justification.

The FDA requires the firm to provide a clear explanation and evidence of corrective actions taken to address these violations. This includes clarifying software version discrepancies, validating device performance with appropriate comparator methods, and ensuring all documentation is complete and accurate. Failure to adequately address these issues could result in further regulatory actions.

Company
Visgeneer, Inc.
Inspection date
February 27, 2025
Product type
Devices
Office
Center for Devices and Radiological Health
People
  • Jerry Doane
  • Courtney H. Lias (Acting Director)
View Company Profile

ID · a93f01ff-45d8-497a-a09d-f61dae020012

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 820.30(g)21 CFR 820.198(a)21 CFR 820.5021 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 806.1021 CFR 862.9(a)21 U.S.C. 381(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Visgeneer, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.