WARNING LETTER
•Xiantao Tongda Non-woven Products Co., Ltd.•July 26, 2015

Xiantao Tongda Non-woven Products Co., Ltd. – FDA Warning Letter, Jul 26, 2015

Record Details

An FDA inspection of Xiantao Tongda Non-Woven Products Co., Ltd. in Xiantao, China, from July 25-26, 2015, revealed that the firm's sterile surgical gowns and masks are adulterated because their manufacturing methods, facilities, or controls do not conform to the Quality System (QS) regulation (21 CFR Part 820). Key violations include: 1) Failure to establish and maintain procedures for controlling environmental conditions (21 CFR 820.70(c)), with observations of trash, debris, holes in walls/doors, and soiled materials. 2) Failure to establish and maintain procedures for controlling storage areas to prevent contamination and damage (21 CFR 820.150(a)), noting materials stored outside, soiled components, and unclean bins. 3) Failure to establish and maintain procedures to prevent product contamination by substances (21 CFR 820.70(e)), with employees not following gowning/hygiene requirements. 4) Failure to adequately control nonconforming product (21 CFR 820.90(a)), as defective masks were discarded without proper labeling. 5) Failure to establish and maintain equipment calibration procedures (21 CFR 820.72(a)), with unlabeled calibration status. Additionally, the firm failed to fulfill annual establishment registration and device listing requirements for fiscal year 2015, rendering devices misbranded under section 510 of the Act. The firm's responses to the FDA 483 were deemed inadequate, lacking sufficient detail on recurrence prevention, verification, and implementation documentation. Consequently, devices are subject to refusal of admission into the U.S. The firm must provide a written response within fifteen business days detailing corrective actions, prevention plans, and a timetable, with documentation.

Company
Xiantao Tongda Non-woven Products Co., Ltd.
Inspection date
July 26, 2015
Product type
Devices
Office
Center for Devices and Radiological Health
Person
  • Sean M. Boyd (Office Director)
View Company Profile

ID · 8da46934-0053-4169-b71d-9cb6ad75257d

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.70(c)21 CFR 820.150(a)21 CFR 820.70(e)21 CFR 820.90(a)21 CFR 820.72(a)21 U.S.C. 36021 U.S.C. 360(p)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Xiantao Tongda Non-woven Products Co., Ltd.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.