FDA WARNING_LETTER - Zeppessis Reprocessing, LLC - April 11, 2013
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On April 8-11, 2013, an FDA inspection of Zeppessis in Garden City, Idaho, revealed that the firm reprocesses Stryker Neptune 1 and 2 Waste Management System manifolds, which are considered adulterated devices under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations include: - **Process Validation (21 CFR 820.75(a)):** Failure to validate the 7-step cleaning protocol, the use of industrial degreaser and Cidex OPA-C for manual reprocessing, and the addition of a non-original rubber o-ring. The firm's response regarding functional validation of the reflux prevention valve and new cleaning products was inadequate. - **Complaint Files (21 CFR 820.198(a)):** Failure to document a complaint about an oversized o-ring causing a manifold to jam. The submitted procedure lacked required information, including reportability evaluation and device identification. - **Corrective and Preventive Action (CAPA) (21 CFR 820.100(a)):** Failure to investigate a Bacillus count exceeding specifications, identify corrective actions, or verify effectiveness. The CAPA procedure was incomplete. - **Production Process Control (21 CFR 820
- Office
- Seattle District Office
- Person
ID · 9b151527-3004-4579-9fe0-0f2e5d7bcf25
Full citation text and observation details available on the Dashboard.