FDA 483 - Baxter Healthcare Corporation - June 09, 2022
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of Baxter Healthcare Corporation revealed that procedures for corrective and preventive action (CAPA) were not adequately established. The firm failed to initiate necessary software updates for marketed Voalte Nurse Call software to address a design requirement non-adherence, despite an informal plan and a related nonconformance evaluation. This indicates a significant lapse in their quality system regarding product corrections and preventive measures.