Baxter Healthcare CorporationFDA 483 - Baxter Healthcare Corporation - August 18, 2023
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
An FDA inspection of Baxter Healthcare Corporation in Round Lake, IL, identified that procedures for design validation were not adequately established. Specifically, a claim regarding the Novum IQ Syringe Pump's Smart Battery Pack degradation was not validated using production-equivalent units. This indicates a deficiency in ensuring product claims are thoroughly tested according to established procedures.