FDA 483 - Boston Scientific Corporation - November 18, 2021
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of Boston Scientific Corporation's nonclinical laboratory in Saint Paul, MN, revealed multiple deficiencies in adherence to Good Laboratory Practice (GLP) regulations. Observations included inadequate protocols for tissue evaluation during necropsy, failure to accurately record experimental data, and improper retention of wet tissue specimens. Additionally, the firm was cited for unlabeled reagents and a quality assurance unit that did not ensure proper specimen identification, impacting the integrity and reliability of safety studies for implanted medical devices.
ID · 45e3f6b9-3749-4a2d-afc5-51e473960984