FDA 483 - Dynalabs LLC - October 29, 2012
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The FDA Form 483 details significant deficiencies in a facility's quality control and manufacturing processes.
**Sterility Testing and Procedures:** * The firm issues Certificates of Analysis (CoAs) citing USP <71> for sterility testing but does not follow its requirements for the number of articles to be tested based on batch size. * Batch size information is not required from clients, nor is there documentation to ensure clients submit the correct number of articles for initial or re-testing. * Specifically, for drug product glutathione/vitamin C/DMOS, lot (b)(4), the required number of articles per USP <71> was not submitted for testing or re-testing. * Procedure MIC-SOP-0016 for re-testing is inadequate, lacking definitions for "proven lab error" and documentation requirements for invalidation and re-testing of initial out-of-specification (OOS) results.
**Laboratory Controls and Potency Analysis:** * No written procedure exists for potency analysis of methylprednisolone. * Sample calculations for potency analysis lack descriptions in analytical worksheets or written procedures, including drug product bug volume corrections. This was observed in multiple investigation reports for various drug products. * Sample preparation descriptions in test schedules are insufficient, lacking details on sample handling, equipment type (e.g., autopipette), glassware, and protection from degradation (e.g., for methylpred
- Company
- Dynalabs LLC
- Inspection Date
- October 29, 2012
ID · 58279535-db6a-49e3-9a5b-9a3909f2a97c
Full citation text and observation details available on the Dashboard.