FDA 483 - E-Beam Services, Inc. - January 27, 2020
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of E-Beam Services, Inc. in Cranbury, NJ, a sterilizer facility, revealed significant deficiencies in equipment maintenance documentation and adherence to calibration procedures. The firm failed to log repairs and associated activities for non-conformances related to their E-Beam sterilizer and did not follow established procedures for line speed calibration, without documenting any rationale for the deviation. These findings indicate a lack of control over critical equipment processes.
ID · 2173bd9b-0398-4a36-80a8-b06ddd9a2c5e