FDA 483 - Genentech, Inc. - September 27, 2011
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
This FDA Form 483 was issued to Genentech Inc., a manufacturer located at 1 DNA Way, South San Francisco, CA, following an inspection from September 2 to September 27, 2011. The report was addressed to Andrea J. Goddard, Senior Director and SSF Site Quality Head.
The inspection revealed four observations:
1. **Inadequate Investigation Records (Repeat Observation):** Written records of investigations into discrepancies and batch failures lacked conclusions and follow-up. * Deviation Report #506273 (disintegrating gaskets on vial washer) failed to include comparison of particulates from culled Avastin vials with the gasket material to rule out contamination of a quarantined lot. * Deviation Report #464179 (capper power outage) failed to describe actions taken for a capper oil leak and disposition of in-process vials, lacking assurance of repair before resuming manufacturing or proper discard of vials.
2. **Inappropriate Equipment Design:** Equipment used in manufacturing was not of appropriate design for its intended use. * No data assures Avastin vial conveyance and capping occurs in a "Controlled Primary HEPA filtered Area" as described in BLA #125085. A gap in the capping area's HEPA filter coverage appeared to allow influx of Class [redacted] air. * Lack of adequate documented evidence to
- Company
- Genentech, Inc.
- Inspection Date
- September 27, 2011
ID · 18f90c1b-4f69-4c48-99f7-a22fc019324c
Full citation text and observation details available on the Dashboard.