FDA 483 - inx Medical, LLC - October 27, 2022
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of inx Medical, LLC, a manufacturer of the Nexus Ligator medical device, revealed significant deficiencies in their quality system. The firm failed to maintain an adequate design history file, lacking documentation for design reviews, verification, validation, and labeling changes. Additionally, the company's corrective and preventive action procedures were found to be inadequate, with instances where CAPAs were not initiated for non-conformances and effectiveness checks were not performed.
ID · d009c801-0f81-4248-b07d-242e950faa60